5.1 Analyzer Module Setup

5.1.1 Calibration Parameters – Span gases

Main Menu — Analyzer and I/O-module expert configuration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module setup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Span gases

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MLT25/CH2/R2

 

 

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

-- Span Gas Definition --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Actual span gas concentration:

 

 

 

250.0

ppm

 

 

 

 

 

 

 

 

Spangas range -1:

 

 

 

 

 

 

50.00

ppm

 

 

 

 

 

 

 

 

Spangas range -2:

 

 

 

 

 

 

250.0

ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Spangas range -3:

 

 

 

 

 

 

2500

ppm

 

 

 

 

 

 

 

 

Spangas range -4:

 

 

 

 

 

 

10000

ppm

 

 

 

 

 

 

 

 

Span gas unit:

 

 

 

 

 

 

 

 

 

ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

"ppm" --> "mg/Nm3" conversion factor:

 

 

 

1.000

 

 

 

 

 

 

 

 

Concentration in span gas units:

 

 

 

95.00

ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

 

 

 

 

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

F3

 

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Span Gas Definition" you can set up the default value for each range of the current channel, the actual span gas concentration and the desired span gas unit. The concentration of each range should be a value between 70 and 110% of the end of range value.

Set up parameters:

¨Select any line of variables with the -key or the -key.

¨Select the variable with the 8 -key or the -key.

¨Select any digit with the -key or the -key and adjust a new value with the -key or the -key

resp. select the whole parameter with the -key or the -key.

¨Confirm the new value with the 8 -key or

cancel and go back to the last value with the F2 key.

The last two lines of variables:

These lines are only information lines. The variables cannot be changed.

Note:

The "ppm ® "mg/Nm3" conversion factor can be set up in the menu "General Concentration Measurement Setup" (see section 5.1.10 p. 5-54)!

5 - 6

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98