Lines of variables "Max. zero calibration deviation" / "Max. span calibration deviation":
The deviation between two zero or span gas concentrations will be determined during the stability time of the calibration procedure (see menu "Calibration Procedure Setup"
p.
If you choose a very small deviation tolerance, perhaps the calculated deviation is out of range during the whole calibration procedure and the calibration will run into time out. If you choose a very large deviation tolerance, the calibration will also be done in ranges with a little stability.
Value is usually: 10% or 20%
Line of variables "Check deviation":
♦Enabled: The stability and tolerance control during the calibration procedure will be based on the values of the menu "Tolerances".
♦Disabled: The calibration procedure is running without stability and tolerance control. Any Calibration will be accepted.
Example: |
|
♦ End of range: | 1000 ppm |
♦ Max. calibration deviation: | 20% (equivalent 200 ppm) |
♦ Desired value: | 990 ppm |
♦ Display: | 720 ppm |
♦ Deviation: | 250 ppm (more than 200 ppm!) |
♦ Result: | Max. calibration procedure time over limit; |
| Calibration will be canceled ! |
Potential solutions:
♦Adjust 100% in the line "Max. zero calibration deviation" resp. "Max. span calibration deviation"
or
select "Disabled" in the line "Check deviation".
♦Result: Any Calibration will be accepted!
5 - 8 | NGA 2000 |
90003482(2)