5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Time controlled calibration
Main Menu — Analyzer and
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Analyzer module setup | ||||||||||||||||
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Calibration parameters | ||||||||||||||||
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Time controlled calibration | ||||||||||||||||
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| TAG |
|
|
|
|
|
|
|
| 37.50 ppm |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| Zero calibration - Day of week: |
|
|
|
|
| Never |
|
|
|
|
|
|
| |||||||
| - Hour: |
|
|
|
|
|
|
| 0 |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| - Minute: |
|
|
|
|
|
|
| 0 |
|
|
|
|
|
|
| |||||
| Span calibration - Day of week: |
|
|
|
|
| Never |
|
|
|
|
|
|
| |||||||
| - Hour: |
|
|
|
|
|
|
| 0 |
|
|
|
|
|
|
| |||||
| - Minute: |
|
|
|
|
|
|
| 0 |
|
|
|
|
|
|
| |||||
| Zero+Span calibration - Day of week: |
|
| Never |
|
|
|
|
|
|
| ||||||||||
| - Hour: |
|
|
|
|
|
|
| 0 |
|
|
|
|
|
|
| |||||
| - Minute: |
|
|
|
|
|
|
| 0 |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Measure |
|
|
| Channel |
| Back... |
| ShowList... |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 | F2 |
| F3 |
| F4 | F5 |
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Time Control" you can determine the starting time of the zero gas calibration resp. the span gas calibration for one channel. If you want to determine the values of the other channels of an MLT analyzer or analyzer module, press the F3 key.
Notes:
¨The time controlled calibration is only possible if the span gas comes from a valve gear! If the valve gear is not available you have to select "Never" in all lines containing "Day of week".
¨A zero gas calibration should be done before starting a span gas calibration. Therefore the starting time of the zero gas calibration should be different from the starting time of the span gas calibration. If you select the option "Zero+Span calibration" the zeroing will run before the spanning, automatically.
Set up time parameters:
¨Select any line of variables with the ↓
¨Select the variable with the 8
¨Select any digit with the ←
the −
for hour: | 0, 1, 2, 3, ..., 23 |
for minute: | 0, 1, 2, 3, ..., 59 |
¨ Confirm the new value with the 8