Further Parameters:
Via the F5 key (ShowList) you can change to the following submenu:
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| TAG |
|
|
|
|
|
| 37.50 ppm |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| Position in |
|
|
|
| 1 |
|
|
|
|
|
|
| ||||||
| Erase current position from list ! |
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| Erase all positions from list ! |
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| Channel tag: |
|
|
|
| - |
|
|
|
|
|
|
| ||||||
| Procedure type: |
|
|
|
| - |
|
|
|
|
|
|
| ||||||
| Interval mode: |
|
|
|
|
| Never |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Start time: |
|
|
|
|
|
| - |
|
|
|
|
|
|
| ||||
| Start date: |
|
|
|
|
|
| - |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Measure |
|
|
|
|
| Back... |
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 | F2 | F3 | F4 | F5 |
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In this submenu you can determine some further parameters of the time controlled calibration.
Lines "Position in
If you have disabled a time controlled calibration in the menu "Time Control" you have to reset the memory to avoid further calibrations by time control. Starting the function "Erase current position from list !" will reset the memory for the position selected in the line "Position in
There are three positions belonging to the following lines in the menu "Time Control":
Position 1: Zero calibration
Position 2: Span calibration
Position 3: Zero+Span calibration
Starting the function "Erase all positions from list !" will delete the
The last five lines of variables:
These lines will display the parameters of the position selected in the line "Position in
90003482(2)
NGA 2000 | 5 - 13 |