From the menu "Analyzer Module Setup and Controls" you can change to all menus and submenus of the expert configuration level via the corresponding menu lines.

Changing to the menus and submenus:

¨Select the menu line you want with the -key or the -key.

If necessary: Change to a further menu page with the F5 key (More...)

¨Change to the corresponding menu/submenu with the 8 -key or the -key.

Menu "Load Factory Configuration":

If you press the F2 key (ManData) in the menu "Analyzer Module Setup and Controls (1/3)", you will change to the following menu:

TAG

95.00 ppm

-- Load Factory Configuration --

- BE CAREFUL with this function -

Replace current configuration with factory settings !

Measure

 

 

Back...

 

F1

F2

F3

F4

F5

With the function of this menu you can delete the RAM data and load the factory settings from the Flash-EPROM.

Caution:

This function is irreversible after starting (and confirmation) ! Your RAM settings will be deleted !

Note:

The same function is available in the menu "Load/Save Analyzer Module Configuration" ! (see section 5.1.17 p. 5-64)

Starting the function:

¨Press the 8 -key or the -key or

cancel and go back to the last menu page with the F4 key (Back...) or the -key.

¨If asked: Confirm with the F2 key (Yes) or

cancel and go back to the last menu page with the F4 key (Back...) or the -key.

Further Ramifications via softkeys:

 

Change to the single component display of the current channel !

F1

 

Change to the other available channels

F3

 

of an MLT analyzer or analyzer module !

+

+

Chapter 4 !

See channel tag !

5 - 4