Structure of chapter five:

At the begin of any chapter you will find the way to a certain submenu of the MLT software starting from the line "Analyzer and I/O-module expert configuration" in the "Main Menu". The way will be described with the software catchwords, that you have to enter one after another to reach the corresponding submenu. At the end of the catchword listing you will find the illustration of the LCD screen. After that you can read the set up instructions and explanations. Sometimes the illustration will be completed by sketches or further menu pictures.

Example: You will set up the calibration parameters for the calibration procedure of the analyzer module.

Main Menu — Analyzer and I/O-module expert configuration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module setup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration procedure setup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

 

 

 

-- Calibration Procedure Setup --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Purge time:

 

 

 

 

 

 

 

10

s

 

 

 

 

 

 

 

 

Procedure times-out after:

 

 

 

120

s

 

 

 

 

 

 

 

 

Analog output during calibration:

 

 

 

 

Tracking

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Span ranges:

 

 

 

 

 

 

Together

 

 

 

 

 

 

 

 

Valve position:

 

 

 

 

 

 

Samplegas

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Stability time:

 

 

 

 

 

 

30

s

 

 

 

 

 

 

 

 

Averaging time:

 

 

 

 

 

 

5

s

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

 

Channel

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

 

F3

 

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Calibration Procedure Setup" you can set up the parameters for the cali-

bration procedure of the zeroing and spanning ...

Further explanations and instructions will follow !

5 - 2

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98