5.5 Expert Analyzer Controls
Main Menu — Expert controls and set up
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Expert analyzer controls | |||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| |||
|
|
|
|
|
| Expert controls |
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| Measurement range number: |
|
|
| 3 |
|
|
|
|
|
|
|
| |||||||||
| Range upper limit: |
|
|
|
|
|
| 100.0 ppm |
|
|
|
|
|
|
|
| ||||||
| Range settings... |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Range and functional control: |
|
|
|
| AUTO |
|
|
|
|
|
|
| |||||||||
| Measurement mode: |
|
|
|
|
|
|
|
| NOx |
|
|
|
|
|
|
| |||||
| Ozonator status: |
|
|
|
|
|
|
|
| ON |
|
|
|
|
|
|
| |||||
| Ozonator power: |
|
|
|
|
|
|
| Enabled |
|
|
|
|
|
|
| ||||||
| Zero/Span calibration... |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Ranges with valid calibration: |
|
|
| 3&4 |
|
|
|
|
|
|
|
| |||||||||
| Physical Measurements... |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
| CAL |
| NO/NOx |
| INFO |
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 |
| F2 |
| F3 |
| F4 |
| F5 |
| ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Expert Controls" you can set up several measurement parameters of the analyzer as range or gas selection. Besides, you can change to further menus, where you can start calibrations for instance.
Set up parameters:
¨Select the line of variables or the menu line with the ↓
¨Select the variable or change to the submenu with the ENTER key or the →
¨Select the parameter with the −
¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.
Lines of variables "Measurement range number", "Range upper limit" and menu line "Range settings...":
You can set up the current range for the analyzer in the line "Measurement range number". A change in this line will change the end of range value automatically in the line "Range upper limit". You cannot change the end of range value there. To change it you have to go to the menu "Range Settings" via the corresponding menu line. This procedure will be described on page
Line of variables "Range and functional control":
¨LOCAL:
The platform will be controlled via the display.
90003749(1)
NGA 2000 | 5 - 53 |