| CLD |
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| ||||
|
|
|
|
|
| Units control |
|
|
|
|
|
|
|
|
|
|
|
| ||
| This control only acts on analyzers prior to version 2.2 |
|
|
|
|
|
|
| ||||||||||||
| which cannot set their own units. |
|
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| ppm to vol % switch point: |
| 10000 ppm |
|
|
|
|
|
|
|
| |||||||||
| PPB to PPM switch point: |
| 100.0 ppb |
|
|
|
|
|
|
|
| |||||||||
| Hysteresis: |
|
|
|
|
|
| 10 % |
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
|
|
|
|
| INFO |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| F1 | F2 | F3 | F4 |
| F5 |
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Units control" you can determine the switch point for the signal value from ppb to ppm, respectively from ppm to vol%. But, the
Set up parameters:
¨Select any line of variables with the ↓
¨Select the variable with the ENTER key or the →
¨Select any digit with the ←
¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.
Lines of variables "ppm to vol% switch point" and "PPB to PPM switch point":
In these lines you can determine which signal value shall switch the displayed concentration from ppb to ppm resp. from ppm to vol% (or reverse).
Line of variables "Hysteresis":
The value selected in this line will determine the range for the switching of the concentration unit.
Example:
¨Hysteresis is 10% and switch point ppm
¨The concentration value displayed will change with increasing signal values from ppm to vol% at: 11000 ppm, i.e. the concentration of 11000 ppm will be displayed as 1.1%.
¨The concentration value displayed will change with decreasing signal values from vol% to ppm at: 0.9%, i.e. the concentration 0.9% will be displayed as 9000 ppm.