CLD |
|
|
| 95.0 ppm |
| Select IO modules |
| ||
Select the modules you wish to bind to the current | ||||
analyzer. |
|
| I/O module with 3 alarms | |
|
|
| ||
|
|
|
| Autocal I/O |
|
|
|
| Syscal I/O |
HOME | ESCAPE | MORE | BACK | INFO |
F1 | F2 | F3 | F4 | F5 |
¨Change to the line you want with the −
If there exist more than eight unbound I/O modules (e.g. Autocal I/O) you can change to another menu page with the F3 key (MORE).
¨Press the ENTER key or the →
The tag of the I/O module selected will appear in the line "Proposed bind".
3) Binding:
¨Press the ENTER key or the →
The control module will change to the initializing mode and the binding will be done, automatically.
¨Repeat the steps 1) to 3) to bind further modules you want.
4) Check bindings:
¨Press the ENTER key or the →
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| ||
|
|
|
|
|
| Modules Bound |
|
|
|
|
|
|
|
|
|
|
| |||
| Analyzer module selected: |
|
|
| CLD |
|
|
|
|
|
|
| ||||||||
| Auxiliary module: |
|
|
|
|
| Syscal I/O |
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Auxiliary module: |
|
|
|
|
|
| NONE |
|
|
|
|
|
|
| |||||
| Auxiliary module: |
|
|
|
|
|
| NONE |
|
|
|
|
|
|
| |||||
| Auxiliary module: |
|
|
|
|
|
| NONE |
|
|
|
|
|
|
| |||||
| Auxiliary module: |
|
|
|
|
|
| NONE |
|
|
|
|
|
|
| |||||
| Auxiliary module: |
|
|
|
|
|
| NONE |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| Proposed new bind: |
|
|
|
|
|
| NONE |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
| NEXT |
|
|
| INFO |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| F1 |
| F2 |
| F3 | F4 |
| F5 |
| |||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
90003749(1)
NGA 2000 | 6 - 13 |