6.1.5 System set up

System Reset

Main Menu — Technical level configuration

System set up

System Reset

CLD

95.0 ppm

System Reset

Are you sure?

Re-initializing will destroy all the binds!!!

System reset!

Re-initialize network!

HOME

ESCAPE

 

 

INFO

F1

F2

F3

F4

F5

You can start the following functions in the menu "System Reset":

1) System Reset:

¨The MLT analyzer or the platform will change to the initializing mode automatically, if you press the ENTER key or the -key in the line "System reset!":

This reboot is analogous to the "switching on" of the control module.

2) Re-initialize network:

¨Change to the line "Re-initialize network!" with the -key or the -key.

¨Press the ENTER key or the -key to start the function.

Caution: With the function "Re-initialize network!", you will delete all bindings between analyzer modules and I/O modules and all configuration data of the SIO and DIO modules will disappear!

6 - 14

NGA 2000

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98