Lines of variables "Hardware revision number" / "Software revision number":

In these information lines you will find the hardware revision number and the software revision number of the CLD analyzer module. You cannot select these lines.

Lines of variables "Minimum range" / "Maximum range":

In these lines you can set up which begin of range and which end of range shall be the limit for all ranges. That means: No begin of range in the menu "Range Settings" (chapter

5.1.6p. 5-16or 5.5.1 p. 5-55)can be minor as the value of the line "Minimum range" and no end of range in the menu "Range Settings" can be major as the value of the line "Maximum range".

Options:

Minimum range

0 to 1000 ppm

 

Maximum range

0 to 10000 ppm

Line of variables "Capillary":

In this line you can set up the capillary flow: 50 to 1000 ml/min.

Caution:

If you change the values in the lines "Minimum range", "Maximum range" and "Capillary" you will change measurement procedure of the CLD !

That means:

The amplifying steps will be preselected by these parameters automatically.

¨Wrong parameters in these lines will produce error measurements automatically !

Therefore, you should not change the setups in these lines !

6 - 20

NGA 2000

90003749X(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98