Lines of variables "Hardware revision number" / "Software revision number":
In these information lines you will find the hardware revision number and the software revision number of the CLD analyzer module. You cannot select these lines.
Lines of variables "Minimum range" / "Maximum range":
In these lines you can set up which begin of range and which end of range shall be the limit for all ranges. That means: No begin of range in the menu "Range Settings" (chapter
5.1.6p.
Options: | Minimum range | 0 to 1000 ppm |
| Maximum range | 0 to 10000 ppm |
Line of variables "Capillary":
In this line you can set up the capillary flow: 50 to 1000 ml/min.
Caution:
♦If you change the values in the lines "Minimum range", "Maximum range" and "Capillary" you will change measurement procedure of the CLD !
That means:
The amplifying steps will be preselected by these parameters automatically.
¨Wrong parameters in these lines will produce error measurements automatically !
♦Therefore, you should not change the setups in these lines !