6.3.2.2 Diagnostic Menus

Primary Variable Parameters

Main Menu — Technical level configuration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Diagnostic menus

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module diagnostics

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Primary variable parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

Primary variable parameters

 

 

 

 

 

 

 

 

 

 

 

 

Raw measurement signal:

 

 

 

521350

 

 

 

 

 

 

 

 

 

Signal gain setting:

 

 

 

 

 

32

 

 

 

 

 

 

 

 

 

Current range:

 

 

 

 

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Barometric pressure compensation:

 

Enabled

 

 

 

 

 

 

 

 

 

Pk-pk noise:

 

 

 

 

 

 

 

 

351 ppm

 

 

 

 

 

 

 

 

 

Peltier device voltage:

 

 

 

 

 

 

Normal

 

 

 

 

 

 

 

 

Reference duty cycle:

 

 

 

 

 

0.2785000

 

 

 

 

 

 

 

 

 

Duty cycle coefficient:

 

 

 

 

 

-0.0681700

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(Temperature compensation coefficients.)

 

 

 

 

 

 

 

 

 

 

 

 

Calibration factors...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

 

 

 

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Primary variable parameters" you will find the parameters that the CLD analyzer module will use to calculate its primary variables internal. Besides, you can change to the submenu "Calibration Factors", where you can set up manual the factors for zero/span calibration of the ranges.

Set up parameters:

¨Select the line of variables you want or the menu line "Calibration factors..." with the -key or the -key.

¨Select the variable or change to the submenu "Calibration Factors" with the ENTER key or the -key.

¨Select any digit with the -key or the -key and adjust a new value with the -key or the -key

resp. select the whole parameter with the -key or the -key.

¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.

Line of variables "Barometric pressure compensation":

Barometric pressure compensation may be enabled if another analyzer has a pressure measurement device, and is set to report its reading.

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

NGA 2000

6 - 29