Set up parameters:
¨Press the ENTER key or the →
Note: The zero offset is about 32000 counts for the CLD software revision 2.2.1 (16 bit D/A transformer) and about 520000 counts for the revision 2.3 (20 bit D/A transformer).
¨Confirm your setup with the ENTER key.
¨Go to the line "Span factor" with the ↓
¨Press the ENTER key or the →
¨Confirm your setup with the ENTER key.
Notes:
¨You have to set up the factors for all ranges.
¨With the parameter "Zero offset" you will fix the zero value. The changing of the zero offset will influence the signal immediately (look at the headline).
¨With the parameter "Span factor" you will fix the signal at the span gas setpoint. The changing of the span factor will influence the signal immediately (look at the headline).
¨Conditions for the manual input of factors:
·"Disabled" has to be set up in the line "Calibration adjustment limits" of the menu "Zero/span calibration". Look at page
·"SEPARATELY" has to be set up in the lines "Zero ranges" and "Span ranges" of the menu "Calibration Parameters" (see 5.1.2, p.
¨Attention:
After the calibration with factors you have to calibrate with zero gas and span gas in the usual way!
Saving the factors:
¨You can store the actual zero offset and span factor with the F2 key (STORE) of the menu "Range 1/2/3/4 Factors". You will find the stored settings on the second menu page (see illustration on the next page).
¨You can change to the second menu page of the menu "Range 1/2/3/4 Factors" with the F4 key (HISTORY):