X K 3 1 9 0 - A 1+P
communication mode parameters.
(2) The sequence of parameters setting as following:
Connecting the loadcell , the indicator enters into working mode which is shown in the Graph
Insert the setting jumper into
Setup communication parameters as following:
Step | Operation | Display | Explanation | |
1 | Press[CALB] |
|
| Plug in calibration jumper |
2 |
| [E | **] | Don’t change this data and enter next step. |
| Press[Enter] |
|
|
|
3 |
| [dc | **] | Don’t change this data and enter next step. |
| Press[Enter] |
|
|
|
4 |
| [Pon | **] | Don’t change this data and enter next step. |
| Press[Enter] |
|
|
|
5 |
| [F |
| Don’t change this data and enter next step. |
| Press[Enter] | ******] |
| |
6 |
| [H *****] | Don’t change this data and enter next step. | |
| Press[Enter] | [L *****] |
| |
| Press[Enter] | [td **.**] |
| |
|
|
|
| |
7 |
| [A d r **] | Communication address (01~26) | |
| Press[1] | [A d r01 ] | Example: 1 | |
| Press[Enter] |
|
|
|
8 |
| [bt | *] | Baud rate of serial communication (0~4) |
| Press[1] |
|
| which represent baud rate of |
| Press[Enter] | [bt | 1] | 600,1200,2400,4800,9600 |
9 |
| [tF | *] | Serial communication mode: |
| Press[0] | [tF | 0] | 0- Continous transmitting mode, |
| Press[Enter] |
|
| unreceiving |
|
|
|
| 1- command responding mode |
10 |
| Weighing | End of communication parameters setting | |
|
| mode |
|
|
( Please note the explanation. Don’t change other parameters setting carelessly )
4.7Date, Time(1) Under weighing status of the indicator, press [ Date ] key, and date sign lamp is lighted. Indicator display original data at this time. If it is correct, then press