X K 3 1 9 0 - A 1+P
3. Calibration
According to the Graph
1.Connecting calibration jumper to the 15- pin socket on the indicator back panel. (There is a
2.According to table
Table 4 - 1
step | Operation | Display | Explanation | |
1 | press [ CALB] |
|
| after calibration jumper is inserted |
2 |
| [E | ** ] | Enter division value selecting |
| press [1 ] [0 ] | [E | 10 ] | 1/2/5/10/20/50/100/200 |
| press [Enter ] |
|
| Example: 10 |
3 |
| [ dc | * ] | Enter decimal point (0 |
| press [ 0 ] | [ dc | 0 ] | Example: without Decimal point 0 |
| press [ Enter ] |
|
|
|
4 |
| [pon XY ] | Enter Zero range | |
| press [ 2 ] [ 3 ] | [ pon 23 ] | X: zero set range | |
| press [ Enter ] |
|
| 2% 4% 10% 20% 100% of F.S |
|
|
|
| Y: power on Zero set range it |
|
|
|
| Stands for |
|
|
|
| 2% 4% 10% 20% 100% of F.S |
|
|
|
| example 23 |
5 |
| [F ***** ] | Enter overload warning value when | |
| press [ 3 ] [ 0 ] |
|
| inputting F value,and |
| [ 0 ] [ 9 ] [ 0 ] | [ F 30090 ] | [ Enter ] key, the calibration will begin. | |
|
|
|
| directly pressing [ Enter ],will enter into |
|
|
|
| the tenth step. If pressing [ weighting ], |
|
|
|
| Indicator will return to weighing status. |
6 | press [ Enter ] |
|
| Example: 30090 |
7 |
| [ noLoAdn] | Confirming Zero position. At this time | |
|
|
|
| there is no load on the scale.Pressing |
|
|
|
| [Input] when the stable light is on. |