15In calibration or parameters setting mode, pressing [weigh], it will exit from original calibration or setting mode and return to weighing mode.

2.3Connecting Loadcell to Indicator

1. The 9-pin socket is used for the link-up of loadcell, which has been clearly shown in the graph 2-3.

2.The 4-pin shielded cable is used, +S must be short connected with +E, -S and -E. The indicator does not have the function of long distance compensation.

3.Indicator must be reliably connect ed to Loadcell and shielded-cable of

loadcell must be reliably connect ed to underground . If indicator is powered on, the user should not insert or withdraw the plug in order to protect the indicator and loadcell.

4.Sensor and indicator are static sensitive devices; you must adopt anti-static

measures. In order to protect the operator ,indicator, and relevant devices, you should install lightning rod in the thunderstorm frenquently happening area.