X K 3 1 9 0 - A 1+P

6.3Wrong setting information indicating

1

 

Err

13

Wrong setting of the division value .

2

 

 

 

Decimal point must be less than 5, please reset the decimal bit.

 

Err

14

3

 

 

 

Overload warning must be > 100, please reset it.

 

Err

15

 

 

 

 

Enter new data which should be less than 65000.

 

 

Err

17

6.4Wrong connection information indicating

1

Err

P

It means the printer has trouble or is wrongly connected. push any

 

 

 

key to quit.

2

 

 

It means the load cell signal line is wrongly connected, or its signal is

Err

01

 

 

 

negative.

 

1If this scale is under usage , then can be sure : the load cell connecting wires

 

 

had troubles, or load cell has been damaged.

 

2If this scale hasn’t been calibrated, the user should check the load cell’s

 

 

connective wires first. ( Whether the signal line has been negatively

 

 

connected). If the loadcell cable is corretly linked, but the indicator still

 

 

indicates Err01 , the problem can then be solved by short connecting pin 4

 

 

and pin 6.

3

 

 

It means the load cell’s wires are wrongly linked, or the signal value

Err

02

 

 

 

exceeds the A/D converting range.

 

1If this scale is under usage, then can be sure: the load cell’s connective wires

 

 

had troubles, or load cell is damaged.

 

2If this scale hasn’t been calibrated, please check as following:

a)Carefully check the load cell’s connective wires is right or not.

b)Check if load cell is suitable or not: It should satisfy the following terms: the “scale’s dead load + scale’s rated capacity ” must be less than the load cell’s rated capacity.

c)This may happen in case that there are more than one load cell in the weighing system.

d)Short connecting the pin 4 and pin 1 on the socket of the load cell.

6.5Error of components and solving method

1

Err

18

Key board has problems, It will indicates for ten seconds, then

 

 

 

indicator enters weighing mode.

2

 

 

The data is partly lost in RAM. Operator should put in the calibration

Err

20

 

 

 

plug to self-check, and then pull it out.