press. Press <<Exit>>or <<Enter>> to return to weighing mode. If a programmable data needs to be modified, press keys in the order of <<Modify>><<Figures>><<Enter>> to enter new data. Data order is displayed as follows:

Pressing times

Reminding symbols

 

Data meaning

 

1

YXX

printing is not allowed in weighing mode of

 

 

 

division=0.1e

 

2

A/D

Weight A/D internal code

 

3

P09

Display the weight in zero point

4

P10

Display zero point upon powered on

5

P11

Calibrate zero point internal code

6

F

Full Scale

 

 

 

7

E

Check division 0.0150, 12 options in total

8

P14

Displaying ratio

 

 

9

P15

Displaying Non-linearity edited value within

 

 

the scope of –1%Max1%Max, weight unit

10

P16

Initial zeroing range 2100 unit: %Max.

 

 

This parameter determined the difference

 

 

between present load and calibrated zero

 

 

point with 20%Max as

default in first

 

 

zero-setting upon powered on.

11

P17

Zero-tracing switch, 0 is off and 1 is on

12

P18

Weight unit:

 

 

 

 

0=t

1=kg

2=g

3=mg

13

P19

bit number

 

 

 

 

0=600 1=1200 2=2400 3=4800 4=9600BPS

14

P20

Communication mode:

 

 

 

0= constant sending 1= answer style

15

P21

Communication address:

 

 

 

1255

 

 

16

P22

Printer styles:

 

 

 

 

0=without Chinese characters database

 

 

1=with Chinese characters database (only

 

 

the printer with Chinese characters

 

 

database is applicable at present, this para.

 

 

does not work)

 

 

17

P23

The lighting time of the background light

 

 

1255, unit: Second. 0=no automatic off

18

P24

Scoreboard switch:

 

 

 

 

0=on

1=off

 

 

19

P25

RS232 switch

 

 

 

 

0= on

1=off

 

 

20

P26

Displaying divisions 0.0150 12 options

 

 

total

 

 

 

21

P27

Displaying Version No. For example: H2C