| 1.02 |
|
22 | P28 | The biggest added tare limit switch: | |
|
| 1=permit | 2=prohibited |
|
| If the difference between load and preset | |
|
| zero point is above 1.2Max upon powered on, | |
|
| the instrument alarm will ring, regardless of | |
|
| the present zero point. | |
23 | P29 | Automatic zeroing switch upon powered on: | |
|
| 1= permit | 2= prohibited |
24 | P30 | Zeroing range 2~100 unit: %Max. This | |
|
| parameter determine the biggest difference |
allowed between the zero points of zero- setting and initial
±2%Max
Notes: YXX represents the XX times weighing of the Yth category.
P10 represents the 10th parameter, others is the same.
* means the parameter
**mean the parameter can only be rewritten while calibration switch is on.
25.Calibration procedures:
1)Turn on the calibration switch (push leftwards) (hardware)
2)Press <<Shift>><<Calibrate>>to enter into calibration procedures, “PAS” remind you of entering password. Enter 3190 and then press <<Calibrate>>.
3)“F XXXXX” reminds you of entering the Full Scale. XXXXX means the previous Full Scale. Press <<Figure>> to enter the Full Scale and then press <<Calibrate>> again.
4)“E XXX” reminds you of entering division and xxx means previous division. Press <<Figure>> to enter division and then press <<Calibrate>>.
5)“noL xxxxxxx” represents present A/D internal code, reminding you of confirming zero point. Press <<Calibrate>> to confirm after data is stabilized.
6)“Lod 0” remind you of entering standard weight. Press <<Figure>> to enter standard weight and then press <<Calibrate>>
7)XXXXXXX in “Adl XXXXXXX” represents present A/D internal code, reminding you of the added load confirmed. Adding standard weight, press <<Calibrate>> to confirm after data stabilized.
8)XXX of “P15 XXX” represents the previous
9)The calibrated data is written into EEPROM and then restart.
10)Turn off the calibration switch (push rightwards) to finish calibration.
11)Perform checking and then confirm weighing performance.
If a parameter needs not to be modified during calibration process, press <<Calibration>> to bypass. Or, press <<Exit>> to exit the calibration state and the calibration fails. Other keys are invalid.