| can be settled by increasing division |
Err 30 | After the datum is stabilized for 30 minutes, the scale body powered off automatically |
Err 32 | No signal from scale body received, cause by no battery installed or long distance. |
Err 33 | Accumulation serial No. has reached 99, impossible to accumulate again. |
Err 34 | Illegal division code, new division need to be entered |
Err 35 | Accumulation serial No. is 0, deletion is unavailable |
Err 36 | Accumulation condition unconformable, accumulation unavailable |
Err 37 | The data entered from keyboard is beyond the range |
Err 38 | Attempt to modify |
Err 40 | The calibrated switch is off, preset data modification unavailable |
Err 41 | Error password entered |
| Minus ratio: |
Err 42 | Caused by (1) Load cell is connected in wrong direction |
| (2) wrong calibration operation program |
| (3) The change of inner code is below 5000 after adding standard weight |
|
|
Err 43 | The |
Err 50 | Zero point exceeds initial zeroing range upon switching on |
Err 51 | No answer signal received from upper computer during mode transmission |
CHAO | Weight exceeds Max+9e or preset weight limit |
VI Error message from wrong inner operation
Usually, inner operation error is resulted from wrong para.setting, low power, no signal, etc. Please check and revise the wrong parameter. If cannot be solved, please record all datum and then contact the Technical Department of Shanghai Yaohua Weighing System Co. Ltd.
Error Message | Comments | |
XXXX | 2Error | In program address XXXX, exceeding error happens in |
|
| subprogram of multiplication |
XXXX | 3Error | In program address XXXX, the error of being divided by zero |
|
| happens in subprogram of division |
XXXX | 4Error | In program address XXXX, exceeding error happens in |
|
| subprogram of addition and subtraction |
XXXX | 5Error | In program address XXXX, exceeding error happens in |
|
| subprogram of BCD code becoming floating code. |
XXXX | 6Error | In program address XXXX, exceeding error happens in |
|
| subprogram of floating codes becoming BCD codes |
XXXX | 7Error | In program address XXXX, being powered off automatically 3 |
|
| minutes later due to low power |
XXXX | 8Error | In program address XXXX, being powered off automatically 3 |
|
| minutes after no signal received from the scale body. |