CARE AND MAINTENANCE

PROCEDURE

Carry out hygienic preparation of the unit and accessories as described on page 16.. We recommend using a 0..5% TERRALIN solution for wipe disinfection.. Follow the instructions enclosed with the disinfectant..

WARNING!

• Take special care that no liquids enter the unit, as this may cause damage..

• You should under no circumstances use a cleaning agent.. Cleaning agents containing alcohol or grease pose a fire risk in combination with compressed oxygen..

• Pay special attention to the oxygen inlet and outlet to make sure they remain free of dust, etc.. If the oxygen inlet connection becomes contaminated with dirt, oil, or grease, DO NOT USE OR ATTEMPT TO CLEAN. Contact your supplier for service or repair..

CAUTION!

 

• The carrying bag must never be washed in a washing machine, spin-dried or dried in a laundry

 

drier. .

 

* NOTE:

suitable gloves (e..g.. household or disposable gloves) for disinfection

You are advised to use

work. .

carrying bag, be careful not to scrub the plastic window and do not

When cleaning your

 

roughen the seams.. Repeat cleaning, if necessary.. Hang the bag in a well-ventilated area and

 

allow to air dry.. Do not hang in direct sunlight, as this may cause its external fabric to fade..

PATIENT CHANGE

Carry out a wipe disinfection on the unit’s surfaces before you hand the unit over to a new patient..