CARE AND MAINTENANCE

FUNCTIONAL CHECK

CHECKING FOR LEAKS

1.Close the valve on the oxygen cylinder..

2.Depressurize the EVOLUTIONTM MOTION by inhaling several times using the nasal cannula.. The gauge indicator should drop to zero..

3.Check that all screwed unions and tube connections are tight.. If necessary, tighten them by hand. * NOTE: Do not use a tool.

4.Ensure that the unit is set to the “PULSE” position..

5.Slowly open the valve on the oxygen cylinder until the needle in the gauge indicator no longer moves..

6.Close the cylinder valve again..

7.Observe the needle in the gauge indicator for approximately one minute..

• If the needle remains in its position, everything is OK..

• However, if the contents indicator shows a continuous decrease in pressure, there is a leak in the system.. In this event, contact your Home Care Provider..

 