ASSEMBLY AND USE

OPERATING INSTRUCTIONS (Cont.):

The Active setting should be used during exercise or activity and the Rest setting should be used when inactive or at rest.. The Rest and Active settings should be selected in accordance with a physician’s prescription or with a physician’s approval..

The motion sensor will select Rest or Active mode in accordance with the below table..

ACTIVITY

RESULT

Motion occurs for less than 2 seconds

Remain in Rest mode

Motion occurs for more than 2 seconds

Switch to Active mode

If motion occurs between 2 and 4 seconds

Revert to Rest mode 10 seconds after

 

motion ceases

If motion occurs for more than 4 seconds

Revert to Rest mode 50 seconds after

 

motion ceases

*NOTE: To help prevent possible damage to the unit, keep the EVOLUTIONTM MOTION conserver in a carrying bag. Several bags are available for use with different cylinder sizes and configurations.

STEP 6: When finished using the system, turn off the oxygen supply cylinder valve and continue breathing through the nasal cannula until no further oxygen is detected..

STEP 7: Remove the nasal cannula..

STEP 8: When not in use, store in a clean, dry location..