Exergen TAT-5000 manual Care and Maintenance

Models: TAT-5000

1 32
Download 32 pages 55.96 Kb
Page 25
Image 25

Care and Maintenance

Battery: A standard alkaline 9V battery provides approximately 15,000 readings. **

To replace, loosen the single screw at the bottom of the instrument and remove the battery cover. Disconnect the old battery and replace with a new one in the same location. Replace the cover, and tighten the screw. Use only high quality alkaline batteries or equivalent.

Handling: The TemporalScanner is designed and built to industrial durability standards in order to provide long and trouble-free service. However, it is also a high precision optical instrument, and should be accorded the same degree of care in handling as you would provide other precision optical instruments, such as cameras or otoscopes.

Cleaning the case: The TemoralScanner case can be wiped down with any hospital approved disinfectant, including bleach.

Cleaning the sensor lens: With normal use, the only maintenance required is to keep the lens on the end of the probe clean. It is made of special mirror-like, coated silicon infrared-transmitting material. However, dirt, greasy films or moisture on the lens will interfere with the passage of infrared heat and affect the accuracy of the instrument. Regularly clean the lens with a cotton swab dipped in alcohol in accordance with the instruction label on the instrument (see below). Use only light force for cleaning, to avoid damaging the lens. Water can be used to remove any residual film left by the alcohol. Do not use bleach or other cleaning solutions on the sensor lens.

CLEANING INSTRUCTIONS

ON THE TAT-5000

Sterilization: The industrial grade housing and design of the electronic components allow for completely safe disinfecting with any accepted solution.

Calibration: Factory calibration data is installed via a computer which communicates with the TemporalScanner’s microprocessor. The instrument automatically self-calibrates each time it is turned on using this data, and will never require recalibration. If readings are not correct, the instrument should be returned for repair.

Care and Maintenance

23

Page 25
Image 25
Exergen TAT-5000 manual Care and Maintenance