Exergen TAT-5000 manual Verifier Specifications, Repair If repair is required

Models: TAT-5000

1 32
Download 32 pages 55.96 Kb
Page 29
Image 29

Accuracy Limits: Comparison between the reference instrument(s) and the instrument being verified should be within ±0.4 ºF (0.2 ºC) for acceptable limits. If not, repeat the process. In the event they still differ by more than the acceptable limits, call Exergen for repair or replacement of the failed instrument.

Verifier Specifications:

Power Source

9-volt battery, or 9-volt power supply.

 

 

Battery Life

Approximately 1 hour continuous use.

 

 

Low Voltage Indicator

Red LED shuts off when battery voltage drops below

 

~5 volts.

Temperature Range

97-104 ºF (36-40 ºC)

 

 

Cleaning

Wipe down with alcohol or any hospital approved

 

disinfectant. Do not immerse.

 

 

Repair

If repair is required:

Contact Exergen at (617) 923-9900 for a Return Materials Authorization (RMA) Number.

Mark the RMA number on the outside of your package and packing slips.

Include a description of the fault if possible.

Send the instrument freight/postage prepaid to:

Exergen Corporation 400 Pleasant Street Watertown, MA 02472

The instrument will be returned freight/postage prepaid.

Care and Maintenance

27

Page 29
Image 29
Exergen TAT-5000 manual Verifier Specifications, Repair If repair is required