9 - 13

Safety

Designed to Installation Category III per IEC 1010, CSA C22.2 (including approval), and UL1244 (including approval) for 600V measurements on industrial power distributions systems.

MAX. Input voltage on

300V rms, 600V rms with

INPUT A or INPUT B

 

10:1 probe

MAX. Input voltage on EXTernal input

300 V rms

Surge Protection

4 kV, 6 kV with 10:1 probe

per IEC 664 and ANSI/IEEE C62.41 test method

Maximum Isolation to Earth

600V rms from any terminal

Optical Interface

Optically Isolated to 600V rms

ACCESSORY INFORMATION

Passive Probe 10:1 (PM8918)

Attenuation: x10

Useful Bandwidth: 100 MHz

Input resistance: 10 M

Input capacitance: 15 pF

Max. V x Hz: 109

Cable length: 1.5m

Max. signal Voltage: 600V rms

Max. surge Voltage: 6 kV

Max. Voltage to Ground: 600V rms

Double insulated per IEC1010-1 for 600V.

Optically Isolated RS-232 Adapter /Cable (PM9080/001)

Serial Printer Interface

.......................EPSON FX/LQ,

 

Deskjet Thinkjet, Deskjet, Laserjet

 

Print screens and waveforms or

 

log meter and cursor readings

Full PC Interface

.......................Transfer setups,

 

waveforms, measurements, etc.

Capture Screen

With FlukeView™ software

(optional for Fluke 92B, 96B and 99B)