Fluke 99B, 105B manual Using the ScopeRecord function

Models: 99B 105B

1 176
Download 176 pages 12.7 Kb
Page 93
Image 93

Using in Scope Mode

5 - 23

Using the ScopeRecord™ function

Before you start the ScopeRecord, make the following preparations:

1.Determine whether you need a single ScopeRecord that stops when the record memory is full (SINGLE

SWEEP), or a continuous ScopeRecord that stops only

by pressing .

To make this selection, do the following:

(a)From the Min Max Record menu, press (MORE) to open the menu.

(b)Use to highlight TRACE REPEAT; then

press to select.

(c)Highlight the desired option; then press to select.

2.Use to set the appropriate range for the expected signal. During ScopeRecord, the test tool maintains the present range.

3.Use to set the best time base in the range from 20 ms/DIV to 60 s/DIV. During ScopeRecord the test tool maintains the present time base.

Do the following to start a ScopeRecord:

1.From the Min Max Record menu press (START).

2.If there is still data in memory from the previous recording, a message appears asking you to confirm that you want to start recording.

NOTE

Starting a recording overwrites data stored from the previous recording.

Press (YES) to start the ScopeRecord.

Press (NO) to cancel the ScopeRecord function.

3.To stop ScopeRecord, press .

Use the menu on the bottom display to examine the results, as described in the next section starting with step 2.