Specifications
Dual Input Oscilloscope 4
63Horizontal
Scope Modes...................................Normal, Single, Roll
Ranges
Normal:
equivalent sampling (Fluke 123)..... 20 ns to 500 ns/div
equivalent sampling (Fluke 124)..... 10 ns to 500 ns/div
real time sampling................................... 1 µs to 5 s/div
Single (real time)..................................... 1 µs to 5 s/div
Roll (real time)......................................... 1s to 60 s/div
Sampling Rate (for both channels simultaneously)
Equivalent sampling (repetitive signals)........................
up to 1.25 GS/s
Real time sampling:
1 µs to 5 ms/div.......................................... 25 MS/s
10 ms to 60 s/div ..........................................5 MS/s
Time Base Accuracy
Equivalent sampling.................. ±(0.4% +0.04 time/div)
Real time sampling.................... ±(0.1% +0.04 time/div)
Glitch Detection..................40 ns @ 20 ns to 5 ms/div
200 ns @ 10 ms to 60 s/div
Glitch detection is always active.
Horizontal Move ...........................................10 divisions
Trigger point can be positioned anywhere across the
screen.
Trigger
Screen Update.............................. Free Run, On Trigger
Source..............................................................A, B, EXT
EXTernal via optically isolated trigger probe ITP120
(optional accessory)
Sensitivity A and B (Fluke 123)
@ DC to 5 MHz.............................0.5 divisions or 5 mV
@ 25 MHz.................................................. 1.5 divisions
@ 40 MHz..................................................... 4 divisions
Sensitivity A and B (Fluke 124)
@ DC to 5 MHz.............................0.5 divisions or 5 mV
@ 40 MHz.................................................. 1.5 divisions
@ 60 MHz..................................................... 4 divisions
Slope................................................... Positive, Negative
Video on A..........................interlaced video signals only
Modes................................................Lines, Line Select
Standards.......................... PAL, NTSC, PAL+, SECAM
Polarity.............................................. Positive, Negative
Sensitivity .........................................0.6 divisions sync.