53IIB/54IIB Users Manual

General

Weight

280 g (10 oz)

 

 

Dimensions

2.8 cm × 7.8 cm × 16.2 cm

(without

(1.1 in × 3 in × 6.4 in)

holster)

 

Battery

3 AA batteries

 

 

Certification

P, )

Safety

CAN/CSA C22.2 No. 61010-1-04, ANSI/UL

 

61010-1:2004, EN/IEC 61010-1:2001

 

 

EMC

EN/IEC 61326-1:2006

CAT I

OVERVOLTAGE (Installation) CATEGORY

 

I, Pollution Degree 2 per IEC1010-1*

* Refers to the level of Impulse Withstand Voltage protection provided. Category 1 products should not be attached to mains circuits.

80 PK-1 Thermocouple (supplied with thermometer)

Type

Type K, Chromel Alumel, bead style

 

 

Temperature

-40 oC to +260 oC

Range

(-40 oF to +500 oF)

Accuracy

±1.1 oC (±2.0 oF)

Electrical

Measurement

J-type: -210 oC to +1200 oC

Range

(-346 oF to +2192 oF)

 

K-type: -200 oC to +1372 oC

 

(-328 oF to +2501 oF)

 

T-type: -250 oC to +400 oC

 

(-418 oF to +752 oF)

 

E-type: -150 oC to +1000 oC

 

(-238 oF to +1832 oF)

 

N-type: -200 oC to +1300 oC

 

(-328 oF to +2372 oF)

 

R- and S-type: 0 oC to +1767 oC

 

(+32 oF to +3212 oF)

Display

0.1 oC / oF / K <1000o

1.0 oC / oF / K ≥1000o

Resolution

 

 

20