53IIB/54IIB
Changing Setup Options
Changing the Thermocouple Type
1.Press or until the display shows TYPE.
2.Press to display the thermocouple type choices. The currently selected thermocouple blinks.
3.Press or until the thermocouple you want appears on the display.
4.Press to store the thermocouple type in memory.
Changing the Offset
You can adjust the thermometer's readings to compensate for the errors of a specific thermocouple. See "Using the Offset to Adjust for Probe Errors." The allowable adjustment range is ±5.0 oC or K, and ±9.0 oF.
Model 54: You can store individual offsets for T1 and T2.
1.Press or until the display shows and T1 or T2.
2.Press to indicate that you want to change the offset setting.
The temperature measurement plus the offset appears in the primary display. The offset appears in the secondary display.
3.Press or until the primary display shows the correct reading.
4.Press to store the offset setting in memory.
Remember to reset the offset to 0.0 when it is no longer needed. The offset automatically resets to 0.0 when you change the thermocouple type.
11