This section acquaints you with some basic operations of the calibrator.

Proceed as follows to perform a voltage-to-voltage test:

1.Connect the calibrator’s voltage output to its voltage input as shown in Figure 5.

2.Press Oto turn on the calibrator. Press lto select dc voltage (upper display).

3.If necessary, press Mfor SOURCE mode (lower display). The calibrator is still measuring dc voltage, and you can see the active measurements in the upper display.

4.Press lto select dc voltage sourcing.

5.Press Yand Zto select a digit to change. Press X

to select 1 V for the output value. Press and hold J to enter 1 V as the 0 % value.

6.Press Xto increase the output to 5 V. Press and hold G to enter 5 V as the 100 % value.

7.Press H and Ito step between 0 and 100 % in 25 % step increments.

Shut Down Mode

The calibrator comes with the Shut Down mode enabled for a time duration set to 30 minutes (displayed for about 1 second when the calibrator is initially turned on). When the Shut Down mode is enabled, the calibrator will automatically shut down after the time duration has elapsed from the time the last key was pressed. To disable the Shut Down mode, press O and Y simultaneously. To enable the mode, press O and Z simultaneously. To adjust the time duration, press O and Z simultaneously, then press X and/or W to adjust the time between 1 and 30 minutes.