Fluke 724 user manual Ohms Measurement Ohms Source

Models: 724

1 56
Download 56 pages 59.07 Kb
Page 50
Image 50

724

Users Manual

Ohms Measurement

Ohms Source

Ohms Range

0 to 400

400 to 1.5 k

1.5 to 3.2 k

 

Accuracy ± Ω

 

 

 

4-Wire

 

2- and 3-Wire*

 

 

 

0.1

 

0.15

0.5

 

 

 

1.0

1

 

 

1.5

 

 

 

Ohms Range

15 to 400

15 to 400

400 to 1.5 k

1.5 to 3.2 k

Excitation Current from

Measurement Device

0.15to 0.5 mA

0.5to 2 mA

0.05to 0.8 mA

0.05to 0.4 mA

Accuracy

± Ω

0.15

0.1

0.5

1

Excitation Current: 0.2 mA

Maximum input voltage: 30 V

Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ± 0.005 % of range per °C

*2-wire: Does not include lead resistance.

3-wire: Assumes matched leads with a total resistance not exceeding 100 .

Resolution

15 to 400

0.1

400 to 3.2 k

1

Temperature coefficient -10 °C to 18 °C, +28 °C to

55 °C: ± 0.005 % of resistance range per °C