Temperature Calibrator
Using Source Mode
Simulating Thermocouples
Connect the calibrator TC input/output to the instrument under test with thermocouple wire and the appropriate thermocouple
Note
One pin is wider than the other. Do not try to force a miniplug in the wrong polarization. Figure 13 shows this connection.
Proceed as follows to simulate a thermocouple:
1.Attach the thermocouple leads to the appropriate TC miniplug, then to the TC input/output as shown in Figure 13.
2.If necessary, press Mfor SOURCE mode.
3.Press Tfor the TC display. If desired, continue pressing this key to select the desired thermocouple type.
4.Enter the temperature you want by pressing Xand Wkeys. Press Yand Zto select a different digit to edit.
Simulating RTDs
Connect the calibrator to the instrument under test as shown in Figure 14. Proceed as follows to simulate an RTD:
1.If necessary, press Mfor SOURCE mode.
2.Press Rfor the RTD display.
Note
Use the 3W and 4W terminals for measurement only, not for simulation. The calibrator simulates a
3.Enter the temperature you want by pressing Xand Wkeys. Press Yand Zto select a different digit to edit.
4.If the 724 display indicates ExI HI the excitation current from your device under test exceeds the limits of the 724.