Fluke 76-908, 76-907 Preparation for Scanning, Tests with 3D Resolution and Slice 3DRAS Phantom

Models: 76-908 76-907

1 26
Download 26 pages 13.35 Kb
Page 12
Image 12
2.2 Preparation for Scanning

Nuclear Associates 76-908 & 76-907

Operators Manual

2.2 Preparation for Scanning

2.2.1Positioning the Phantom

The 3DRAS phantom can be placed in a head coil or a body coil. The center of the phantom should coincide approximately with the center of the RF coil. The UAL phantom can be used only with a body coil.

2.2.2 Scanning Parameters

For all measurements, scan conditions should be carefully recorded. Scan conditions records should include:

Any image processing which may have been used

Field-of-view or zoom factor

Image matrix size

Imaging coil

Number of signal (excitation) acquisitions

Phantom and phantom material

Pulse sequence name or code and software version number

RF power settings

Scan timing parameters (TE, TI, TR)

Slice excitation order

Slice number and thickness

Tuning parameters

2.3 Tests with 3D Resolution and Slice (3DRAS) Phantom

2.3.1 Resonance Frequency

Resonance frequency is defined as that RF frequency (f) which matches the static B-field (Bo) according to the Larmor equation. For protons, the Larmor frequency is 42.58 MHz/Tesla, e.g., for a 1.5 Tesla system, the resonance frequency should be 63.87 MHz.

Prior to the performance of any imaging protocol, the resonance frequency must be checked first. Changes in the resonance frequency reflect changes in the static magnetic field (B-field). Changes in the B-field may be due to superconductor "run down" (typically on the order of 1 ppm/day, e.g., about 60 Hz/day at 1.5 Tesla), changes in current density due to thermal or mechanical effects, shim-coil changes, or effects due to external ferromagnetic materials.

The effects of off-resonance operation relate primarily to a reduction in image signal-to-noise. Secondary effects are reflected in image linearity due to the summation of the image gradients with the inconsistent static B-field value.

It is recommended that a resonance frequency check be performed prior to any measurements and each time a different phantom is used.

The phantom is positioned in the center of the magnet (with all gradient fields turned off) and the RF frequency is adjusted by controlling the RF synthesizer center frequency to achieve maximum signal. Some resistive systems may also allow adjustment of the magnet current in order to alter the magnetic field strength so as to achieve resonance. Most vendors will provide a specific user protocol for

2-2

Page 12
Image 12
Fluke 76-908, 76-907 Preparation for Scanning, Tests with 3D Resolution and Slice 3DRAS Phantom, Positioning the Phantom