Function

(50 Hz to 400

Hz)

Range

Resolution

Accuracy

0 to 200 μA

0.1 μA

±(2% + 3 counts)

 

 

 

0 to 200 μA

0.1 μA

±(1% + 2 counts)

Burden Voltage

<5 mV/μA

<5 mV/μA

Functionf

E

Overload

Protection1

600V rms

600V rms

Input Impedance (Nominal)

>5 MΩ <100 pF

VCheck and LoZ = >2 kΩ <200 pF (ac coupled)2

>10MΩ <100 pF

VCheck and LoZ = >2 kΩ <200 pF2

f

E600V rms

600V rms

600V rms

600V rms

Common Mode Rejection Ratio (1 kΩ Unbalanced)

>60 dB at dc 50 or 60 Hz

>100 dB at dc, 50 or 60 Hz

Open Circuit Test Voltage

<1.5V dc

2.4-3.0V dc

Normal Mode Rejection

>50 dB at 50 Hz or 60 Hz

Full Scale Voltage

To 4.0 MΩ

40 MΩ

<450 mV dc

<1.5V dc

2.400V dc

J

L600V rms

600V rms

1. 3 x 106 V Hz maximum

Short Circuit Current

<500 μA

0.95 mA (typical)

2.2 kΩ input impedance up to 50V. Impedance increases with input voltage to >300 kΩ at 600V.

MIN MAX Recording Accuracy and Response Time

Specified accuracy of the measurement function ±12 digits in dc for changes >200 ms in duration (±40 digits in ac). Typical 100 ms response to 80%.

Example 1: This would mean ±1.2° when recording temperature.

Example 2: This would mean ±12 μA when recording μA or ±12A if used with a dc amp probe (with a mV input).

Page 14
Image 14
Fluke PN 643657 instruction sheet MIN MAX Recording Accuracy and Response Time