Function
(50 Hz to 400
Hz)
Range | Resolution | Accuracy |
0 to 200 μA | 0.1 μA | ±(2% + 3 counts) |
|
|
|
0 to 200 μA | 0.1 μA | ±(1% + 2 counts) |
Burden Voltage
<5 mV/μA
<5 mV/μA
Functionf
E
Overload
Protection1
600V rms
600V rms
Input Impedance (Nominal)
>5 MΩ <100 pF
V∙Check and LoZ = >2 kΩ <200 pF (ac coupled)2
>10MΩ <100 pF
V∙Check and LoZ = >2 kΩ <200 pF2
f
E600V rms
600V rms
600V rms
600V rms
Common Mode Rejection Ratio (1 kΩ Unbalanced)
>60 dB at dc 50 or 60 Hz
>100 dB at dc, 50 or 60 Hz
Open Circuit Test Voltage
<1.5V dc
Normal Mode Rejection
>50 dB at 50 Hz or 60 Hz
Full Scale Voltage
To 4.0 MΩ | 40 MΩ |
<450 mV dc | <1.5V dc |
2.400V dc
J
L600V rms
600V rms
1. 3 x 106 V Hz maximum
Short Circuit Current
<500 μA
0.95 mA (typical)
2.≅2 kΩ input impedance up to 50V. Impedance increases with input voltage to >300 kΩ at 600V.
MIN MAX Recording Accuracy and Response Time
Specified accuracy of the measurement function ±12 digits in dc for changes >200 ms in duration (±40 digits in ac). Typical 100 ms response to 80%.
Example 1: This would mean ±1.2° when recording temperature.
Example 2: This would mean ±12 μA when recording μA or ±12A if used with a dc amp probe (with a mV input).