22

RPT 301 User Manual

S - Status

This command returns the instrument status.

Command:

S

 

Reply:

STATUS nn

Where nn:

01

= Frequency outside limits

 

02

= Diode voltage outside limits

 

04

= Electronics failure

 

Combination errors: a status 06 represents diode

 

voltage and electronics failure.

Result:

no change

I - Identity

 

This command requests the unit to send items of data which describe the unit identity.

Command:

I

 

 

Reply:

type, range, S/N, cal date

Result:

No change in status or operation

Where:

Type

=

Equipment model (Druck RPT 301)

 

Range

=

Pressure range, lower and upper and

 

 

 

pressure type

 

S/N

=

Transducer serial number

 

Cal date =

Date of last calibration

Command: I,S/PIN, Name Value<CR> format

i.When Name = S, Value = nnnn/yy-mm Reply: S/N = nnn/yy-mm

Result: sets displayed serial number

ii.When Name = H, Value = nnnn Reply: HIGH RANGE = nnn mbar

Result: sets displayed highest operating pressure (mbar)

iii.When Name = L, Value = nnnn Reply: LOWEST RANGE = nnn mbar

Result: sets displayed lowest operating pressure (mbar)

K177 Issue No. 3

Page 30
Image 30
GE RPT 301 user manual Status, Identity, Command I,S/PIN, Name ValueCR format