54
Chapter 6: Maintaining Your Filtered Enclosure
Measure of Line Voltage and Current
Smoke Test to determine proper airflow patterns
Lighting Intensity Test (when appropriate)
Noise Level Test (when appropriate)
Vibration Test (when appropriate)
Fluorescent Light Replacement
1. Disconnect the power.
2. Locate the small light reflector located under the control
panel shown in Figure 6-2.
3. Remove the light reflector support by removing two
Phillips screws on the bottom of the light reflector.
4. Rotate and remove the old fluorescent lamp.
5. Reinstall the new fluorescent lamp and light reflector in
reverse order.
6. Power the unit up and try the new fluorescent lamp.
UV Light Replacement
1. Disconnect the power.
2. Locate the UV lamp located inside the enclosure.
3. Surface decontaminate the old UV light before handling it.
4. Rotate and remove the old UV lamp.
5. Reinstall the new UV lamp.
6. Power the unit up and try the new UV lamp.
Motorized Impeller Replacement
The motorized impeller must be replaced as a complete unit.
When the motorized impeller is replaced, the capacitor may also be
replaced. See Appendix A for Replacement Parts Diagram. See
Figure 6-5 for an isometric view of the motorized impeller plenum
assembly. The HEPA filter rests on top of the motorized impeller
assembly.
1. Wear appropriate personal protective equipment to
decontaminate the filtered enclosure and then unplug from the
electrical outlet.
2. Remove HEPA filter per the HEPA filter removal procedure
outlined in this chapter.
3. Consult the wiring diagram in Appendix C of the manual and
disconnect all the wires of the motorized impeller. Be sure to
connect wires on the new motor in the same way the old motor
was wired.