Lumex® AltaDyne AutoFloatTroubleshooting/Specifications

14.0 Troubleshooting

SYMPTOM:

CORRECTIVE ACTION

 

 

Mattress does not

Make certain the air hoses are not kinked, cracked, or split. Verify the hose couplings

are fully engaged to controller. Verify that all 20 air cells are connected to the air manifold

inflate or is too soft.

within mattress.

 

 

 

Power loss indicator

Power has been disconnected from unit while in an operational mode.

flashing and audible

Restore power, put unit in STANDBY mode, then remove power. Unit must always be

alarm sounding

returned to STANDBY mode before disconnecting power.

 

 

Indicator adjacent to

If in AUTO mode, check the connection of the Active Sensor mattress sensor cord to the

ALARM SILENCE

jack on the side of the unit. Reconnect if necessary.

button is flashing and

Verify hose couplings are fully engaged to controller. Verify that all 20 air cells are

audible alarm is

connected to the air manifold within mattress.

sounding.

 

 

 

15.0 Specifications, Control Unit

Enclosure Dimensions

10" x 12" x 4 1/2", (25 cm x 30 cm x 11 cm)

Weight

9.5 pounds (4.3 kg)

 

 

Power Cord

Detachable 14' , #18 AWG minimum, with ground wire

 

 

Overcurrent Protection

Primary 2.5A, 250V, T

 

 

Input

120 -230 VAC, 50/60Hz, 120V – 0.44A, 230V – 0.33A

 

 

Operating Ambient

60 to 85°F (16 to 29°C)

Temperature Range

 

 

 

 

Class I grounded equipment not suitable for use in the presence of a flammable

 

anesthetic mixture with air or with oxygen or nitrous oxide.

 

 

 

Type BF equipment

Classification

 

UL2601-1, CSA C22.2 NO. 601.1

 

EN 60601-1

 

 

 

IPXO, enclosed equipment without protection

 

against ingress of water.

 

 

 

Continuous operation

 

 

Electromagnetic

Meets EN60601-1-2:2001

Compatibility

(CISPR 11 Classified as Class B, Group 1 ISM equipment)

 

 

 

 

9

Page 9