Lumex® AltaDyne AutoFloatTroubleshooting/Specifications
14.0 Troubleshooting
SYMPTOM: | CORRECTIVE ACTION | |
|
| |
Mattress does not | Make certain the air hoses are not kinked, cracked, or split. Verify the hose couplings | |
are fully engaged to controller. Verify that all 20 air cells are connected to the air manifold | ||
inflate or is too soft. | ||
within mattress. | ||
| ||
|
| |
Power loss indicator | Power has been disconnected from unit while in an operational mode. | |
flashing and audible | Restore power, put unit in STANDBY mode, then remove power. Unit must always be | |
alarm sounding | returned to STANDBY mode before disconnecting power. | |
|
| |
Indicator adjacent to | If in AUTO mode, check the connection of the Active Sensor mattress sensor cord to the | |
ALARM SILENCE | ||
jack on the side of the unit. Reconnect if necessary. | ||
button is flashing and | ||
Verify hose couplings are fully engaged to controller. Verify that all 20 air cells are | ||
audible alarm is | ||
connected to the air manifold within mattress. | ||
sounding. | ||
| ||
|
|
15.0 Specifications, Control Unit
Enclosure Dimensions | 10" x 12" x 4 1/2", (25 cm x 30 cm x 11 cm) | |
Weight | 9.5 pounds (4.3 kg) | |
|
| |
Power Cord | Detachable 14' , #18 AWG minimum, with ground wire | |
|
| |
Overcurrent Protection | Primary 2.5A, 250V, T | |
|
| |
Input | 120 | |
|
| |
Operating Ambient | 60 to 85°F (16 to 29°C) | |
Temperature Range | ||
| ||
|
| |
| Class I grounded equipment not suitable for use in the presence of a flammable | |
| anesthetic mixture with air or with oxygen or nitrous oxide. | |
|
| |
| Type BF equipment | |
Classification |
| |
| EN | |
|
| |
| IPXO, enclosed equipment without protection | |
| against ingress of water. | |
|
| |
| Continuous operation | |
|
| |
Electromagnetic | Meets | |
Compatibility | (CISPR 11 Classified as Class B, Group 1 ISM equipment) | |
|
| |
|
|
9