Lumex® AltaDyne AutoFloat

Operator Control Panel

8.0 Operator Control Panel

4

 

6

 

 

 

 

 

Auto Float

 

 

 

 

 

 

APM Plus

 

 

 

 

 

5

 

 

 

 

 

3.2

3.4

 

 

 

 

 

3.3

 

 

 

 

7

 

 

 

 

 

9

 

 

 

 

 

8

3

3.1

10

2

1

1.ON/STANDBY switches Control Unit between On and STANDBY modes.

2.POWER INDICATOR LED indicates AC power is present to the Control Unit.

3.MODE SELECT Press repeatedly to toggle through operating modes, Auto/Float, Auto/A/P, Manual/Float and Manual/A/P.

3.1AUTO MODE (Sensor cord from mattress must be connected to control unit) Mattress will automatically inflate to optimal level based on information received from active sensors in mattress. The indicator adjacent to the icon, will light while automatic mode is active.

Note: Connecting the mattress sensor cord while the unit is not in automatic mode will cause the Control Unit to enter automatic mode immediately. If Sensor cord from mattress becomes disconnected from control unit while in Auto mode, alarm will sound.

3.2MANUAL MODE Mattress inflation level user determined by Comfort Adjust (-, +) setting. The indicator adjacent to the icon will light while manual mode is active.

3.3FLOAT MODE all cells within mattress will be uniformly inflated when this icon is illuminated.

3.4AP MODE. (Alternating Pressure) Adjacent cells within mattress will be alternately inflated and deflated in this mode. The indicator adjacent to the icon will light while AP mode is active.

Note: AP mode is temporarily suspended in max inflate mode.

4.SOFT (–)decreases the inflation level of the bed while Control Unit is in manual mode only. Pressing once will decrease the setting by one comfort level. Pressing and holding will rapidly change the setting.

5.COMFORT CONTROL INDICATOR displays the current inflation level setpoint.

6.FIRM (+) increases inflation level of the bed while in manual mode only. Pressing once will increase the setting by one comfort level setting (~2.5mmHg). Pressing and holding will rapidly change the setting.

7.MAX INFLATE selects/deselects the max inflate mode (maximum mattress firmness). The indicator adjacent to the button will flash while max inflate mode is active. After 30 minutes, max inflate mode will cancel.

8.LOCKOUT activates or cancels lockout mode. Press and hold for three seconds to activate or cancel mode. The indicator adjacent to the button will light while lockout is active. In lockout mode, only the MAX INFLATE button is operational.

All other buttons are temporarily inactive.

9.ALARM SILENCE silences indefinitely, the audible alarm.

10.POWER FAIL INDICATOR flashes when power supplied from line cord is interrupted while the Control Unit is operating. To avoid an inadvertent power fail alarm, place the unit in STANDBY mode before disconnecting power from Control Unit.

5

Page 5
Image 5
Graham Field 753004, 753003 manual Auto Float, Operator Control Panel