Display and alarm | Display and alarm |
Upon connecting to the power supply, the freezer will be on and the display screen will show the freezer’s internal temperature.
Prompts for the working state of display panel:
“Lock” indicator: if the indicator lights up, it means that all keys are locked to avoid misoperation.
“Alarm” indicator: if the indicator lights up or flashes, it means that a fault occurs. Functions of keys:
Cancel buzz: in the locked or unlocked state, press the “Cancel buzz” key at the time of alarm, the alarm sound will be cancelled and the visual alarm will not be cancelled.
Alarm test: it is used to test normal working state of the alarm system and low battery charge level. In the unlocked state, press this key, all the indicators will light up, the buzzer will give an alarm and all digital tubes display “8”. In the locked state, press this key at the time of alarm, corresponding fault alarm code will display. Please identify the fault causes according to the table below.
Alarm | State |
|
|
|
| Fault | Indication | Buzzer |
|
|
|
| |
|
|
|
|
|
| code |
|
|
|
|
|
|
|
High | When | the | internal | E9 | The | indicator | When | the | freezer | is | |||
temperature | temperature | is | higher |
| flashes |
| turned on for the first | ||||||
alarm | than the set temperature |
|
|
| time, the | buzzer alarm | |||||||
| for | high | temperature |
|
|
| function | is | activated | ||||
| alarm |
|
|
|
|
|
| after 5 | hours | according | |||
Low | When | the | internal | E10 | The | indicator | to the default settings; in | ||||||
temperature | temperature | is | lower |
| flashes |
| normal | operation, | the | ||||
alarm | than the set temperature |
|
|
| buzz alarm is given after | ||||||||
| for | low | temperature |
|
|
| 15min |
|
|
|
| ||
| alarm |
|
|
|
|
|
|
|
|
|
|
| |
/ | The | alarm | the buzz alarm is given | ||||||||||
n alarm | freezer |
|
|
|
| indicator | after 30s |
|
|
| |||
|
|
|
|
|
|
| flashes; | black |
|
|
|
|
|
|
|
|
|
|
|
| screen | and |
|
|
|
|
|
|
|
|
|
|
|
| actual | internal |
|
|
|
|
|
|
|
|
|
|
|
| temperature |
|
|
|
|
| |
|
|
|
|
|
|
| show alternately |
|
|
|
|
| |
Ambient | temperature | E14 | The | indicator |
|
|
|
|
| ||||
ambient | above 38℃ |
|
|
| flashes |
|
|
|
|
|
| ||
temperature |
|
|
|
|
|
|
|
|
|
|
|
|
|
Abnormal | Internal main sensor at | E6 | The | indicator | Buzz alarm |
|
| ||||||
sensor | fault |
|
|
|
|
| flashes |
|
|
|
|
|
|
| Ambient | temperature | E0 | The | indicator |
|
|
|
|
| |||
| sensor at fault |
|
| flashes |
|
|
|
|
|
| |||
Low | Insufficient |
| battery | E5 | The | indicator |
|
|
|
|
| ||
battery | charge | or | the | battery |
| flashes 3 times |
|
|
|
|
| ||
charge | switch is not on |
|
|
|
|
|
|
|
|
| |||
level |
|
|
|
|
|
|
|
|
|
|
|
|
|
Note:
Always turn on the battery switch before starting the freezer. The freezer will enter normal working state and charge the battery upon connecting to an AC power supply meeting its requirements. In case of AC power failure, the battery will energize the display screen for normal display. When the voltage of battery reduces to less than a specified value due to discharge, it will stop energizing the display screen and thus the display screen will not display any longer. When the battery is able to supply electricity normally, to completely cut off the freezer’s power supply, just pull out its AC power cord and turn off its battery switch. Then the display screen will not display any longer.
Time setting for automatic alarm recovery
At the time of alarm, press the “Cancel buzz” key on the display panel, buzz alarm will be cancelled (remote alarm will not be cancelled).
If alarm conditions remain, the buzz alarm will automatically recover after a pause of 30min.
Remote alarm terminals
The remote alarm terminals are located in the compressor compartment at the back of freezer. Alarm signals are sent by these terminals with carrying capacity of DC 30V, 2A. Contact output:
In normal conditions, both ends of connecting wire of remote alarm terminals are open circuit. In alarm conditions, both ends of connecting wire of remote alarm terminals are closed circuit.
To use the remote alarm function, connect the circuit as shown in the figure below. Note: The remote alarm function is limited to three alarm modes - high temperature alarm, low temperature alarm and
Note: If several alarms sound simultaneously, each fault code displays 3s.