BPW-101IB_ENGSP.qxd 10/24/05 11:14 AM Page 33

TROUBLESHOOTING

If any abnormality arises during use, please check the following points:

SYMPTOMS

POSSIBLE

CORRECTION

CAUSES

 

 

 

 

 

Unit does not turn

Batteries have

Replace them

on when the

run down.

with two new AAA

START/STOP

 

alkaline batteries.

button is pushed.

 

 

Battery polarities

Re-insert the

 

 

have been posi-

batteries in the

 

tioned incorrectly.

correct positions.

 

 

 

EE measurement

Cuff has been

Wrap the cuff prop-

error symbol shown

placed incorrectly.

erly so that it is

on display or the

 

positioned correctly.

blood pressure

 

 

Did you talk or

Measure again.

value is displayed

move during

Keep wrist

excessively low

measurement?

steady during

(or high).

 

measurement.

Shaking of the wrist

 

with the cuff on.

 

 

 

 

E1 error symbol

Air circuit

Measure again.

shown on display

abnormality.

 

 

 

 

E2 error symbol

Inflation pressure

Switch the unit off,

shown on display

exceeding 300 mmHg.

then measure again.

 

 

 

E3 error symbol

Data error.

Remove and reload

shown on display

 

the batteries.

 

 

 

Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassem- ble or attempt to repair the unit by yourself. Contact infor- mation for HoMedics Consumer Relations Department can be found on the warranty page (See page 35).

33

SPECIFICATIONS

Power Source:

Two 3V DC (AAA) Alkaline batteries

Measurement

Oscillometric

Method:

 

 

 

Measurement

Pressure: 40~250 millimeters Mercury

(mmHg)

Range:

Pulse: 40~199 beats/minute

 

 

 

Accuracy:

Pressure: ±3 mmHg

Pulse: ±5% of reading

 

 

 

Pressure Sensor:

Semi-conductor

 

 

Inflation:

Pump driven

 

 

Deflation:

Automatic Pressure release valve

Memory Capacity:

60 memories for each User (120 total)

 

 

Auto-shut-off:

1 minute after last button operation

 

 

Operation

Temperature: 50°F~104°F (10°C~40°C)

Environment:

Humidity: 40%~85% RH max.

 

 

Storage

Temperature: 14°F~140°F (-10°C~60°C)

Environment:

Humidity: 10%~90% RH max.

 

 

Weight:

0.30 lbs (without batteries)

 

 

Wrist

5.3”~8.7” (13.5~22cm)

Circumference:

 

Dimensions:

2.6”(L) x 1.3”(W) x 3.1”(H)

67mm(L) x 32mm(W) x 78mm(H)

 

 

 

 

Type BF

 

Device and cuff are designed to

 

provide special protection against

 

electrical shocks.

Note: These specifications are subject to change

without notice.

 

 