Specifications
Operating Conditions – All Models
Parameter | Reference |
| Rated Condition | Operative Limits | Transportation | |||||||
|
| Condition |
|
|
|
|
| and Storage | ||||
|
| (at zero |
|
|
|
|
|
|
|
| ||
|
| static) |
|
|
|
|
|
|
|
| ||
|
| °C | °F |
| °C | °F | °C | °F | °C |
| °F | |
|
|
|
|
|
|
|
|
|
|
| ||
Ambient Temperature | 25 ±1 | 77 ±2 |
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
Process Interface |
| 25 ±1 | 77 ±2 |
|
|
| See Figure 13 |
| ||||
Temperature |
|
|
|
|
|
|
|
|
|
|
|
|
Humidity | %RH | 10 to 55 |
|
| 0 to 100 | 0 to 100 | 0 to 100 |
| ||||
Maximum Allowable | MAWP is minimum of Body Rating or Seal Rating (See Model Selection Guide for Seal |
| ||||||||||
Working Pressure (MAWP) | MAWP) |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
| Body | MAWP |
|
|
|
|
|
|
|
| STR93D | 750 psig (52 bar) |
|
|
|
| ||||
|
|
| STR94G | 500 psig (35 bar) |
|
|
|
| ||||
Vacuum Region, | Minimum |
|
|
|
|
|
|
|
|
|
|
|
Pressure - mmHg absolute | atmospheric |
|
|
| See Figure 13 |
|
|
|
| |||
inH2O absolute | atmospheric |
|
|
|
|
|
|
|
|
| ||
Supply Voltage, Current, | Voltage Range: 10.8 to 42.4 Vdc at terminals |
|
|
|
| |||||||
and Load Resistance | Current Range: | 3.0 to 21.8 mA |
|
|
|
|
| |||||
|
| Load Resistance: 0 to 1440 ohms (as shown in Figure 14) |
|
|
|