Brookfield Engineering Labs., Inc. Page 52 M/02-212
VIII. Technical data
R/S Rheometer
Dimensions 480 mm x 300 mm x 290 mm
Weight 8 kg
Nominal operating voltage
Power consumption (average)
Power consumption (maximum)
+/- 15 V, 5 V
12 W
22 W
Ambience conditions
Temperature
in operation
out of operation
Relative humidity (not condensable)
in operation
out of operation
10° to 40°C
10° to 45°C
20% to 80%
10% to 90%
Accuracy ± 1.0 % of maximum range value
± 1 digit
Torque range
Mains operation
0.05 to 50 mNm
Torque resolution 0.01 mNm
Speed range 0.7 min-1 to 800 min-1
Angle resolution 0.8 mrad
Temperature range Depending on temperature control device used
-20°C to +180°C
Shear rate range Depending on measuring system used
0.9 s-1 to 4 × 103 s-1
Shear stress range Depending on measuring system used
0.7 Pa to 3.4 × 104 Pa
Viscosity range
The given range is a standard value (not maximum value)
Depending on measuring system used
1 × 10-3 Pas to 3 × 103 Pas
AC-Adaptor
Dimensions 160 mm x 85 mm x 35 mm
Weight 0.5 kg
Power supply
Mains voltage
secondary voltages (Output)
output-current
output power
from 100 up to 240 V AC
5 V, +/- 15 V DC
2 A, 0.9 / -0.2 A
20 W
Frequency range of mains voltage 50 bis 60 Hz
Environment conditions
Temperature
in operation
out of operation
Relative humidity (not condensable)
in operation
out of operation
+10°C to +40°C
+10°C to +45°C
20% to 80%
10% to 90%