Invacare 1095344, 1095341, 1095343, 1095346, 1095340 Contracture Assembly Angle Adjustment Figure

Models: 1095340 1095341 1095343

1 4
Download 4 pages 63.52 Kb
Page 3
Image 3
CONTRACTURE ASSEMBLY ANGLE ADJUSTMENT (FIGURE 4)

CONTRACTURE ASSEMBLY ANGLE ADJUSTMENT (FIGURE 4)

NOTE:RefertotheGENERALWARNINGSandINSTAL- LATION WARNINGS in the SAFETY SUMMARY of this instruction sheet.

1.Remove the mounting screw and locknut used to lock the angle adjustment of the contracture assembly.

2.Lift the contracture support UP/DOWN to achieve the desired angle.

3.Align the mounting hole in the contracture support with one (1) of seven (7) mounting holes on the angle adjustment bracket.

4.Reinstall the mounting screw and locknut used to lock the angle adjustment of the contracture assembly. Securely tighten.

5.If necessary repeat STEPS 1-4 on remaining con- tracture assembly.

Mounting Screw

Angle

Adjustment

Bracket

Mounting

Hole

Locknut

Contracture Support

FIGURE 4- CONTRACTURE ASSEMBLY ANGLE

ADJUSTMENT

USING THE OPTIONAL QUICK RELEASE PIN (FIGURE 5)

NOTE: Refer to the GENERAL WARNINGS and INSTAL- LATION WARNINGS in the SAFETY SUMMARY of this instruction sheet.

NOTE: The Quick Release Pin replaces the mounting screw that is used to lock the angle adjustment of the contracture support.

NOTE:Ifnotinstalledat time of shipment, thequickrelease pin must be installed by a qualified technician only, using a 1/8-inch pop rivet.

1.Press the button on the quick release pin and insert through the contracture support and the angle ad- justment bracket.

2.If necessary adjust the angle of the contracture as- sembly. Refer to CONTRACTURE ASSEMBLY ANGLE ADJUSTMENT in this instruction sheet.

3.Repeat STEPS 1-2 on the remaining contracture as- sembly.

 

Angle Adjustment

 

Bracket

Pop Rivet

 

Button

Contracture

Quick Release Pin

Support

 

FIGURE 5 - USING THE OPTIONAL QUICK

RELEASE PIN

ANGLE ADJUSTMENT OF THE CONTRACTURE FOOTPLATE (FIGURE 6)

NOTE: Refer to the GENERAL WARNINGS and INSTAL- LATION WARNINGS in the SAFETY SUMMARY of this instruction sheet.

1.Loosen, but DO NOT, remove the two (2) flat screws, washers and locknuts that secure contracture footplate to the footplate hinge.

2.Tilt the contracture footplate UP/DOWN to achieve desired angle.

3.Retighten the two (2) flat screws, washers and lock- nuts that secures the contracture foot plate to the footplate hinge.

4.If necessary repeat STEPS 1-3 on remaining con- tracture footplate.

 

Flat Screws

Contracture

 

Assembly

Contracture

 

Footplate

Footplate

 

Hinge

 

NOTE:Mountinghardware

Washers

shown removed for clarity

 

only.

Lock Nuts

FIGURE 6 - ANGLE ADJUSTMENT OF THE

CONTRACTURE FOOTPLATE

3

Page 3
Image 3
Invacare 1095344 Contracture Assembly Angle Adjustment Figure, Angle Adjustment Of The Contracture Footplate Figure