Invacare 6302G user manual 5.2Cleaning and Disinfecting, Warnings, Operation

Models: 6302G

1 32
Download 32 pages 22.7 Kb
Page 23
Image 23
5.2Cleaning and Disinfecting

5 OPERATION

Perform these procedures in the following order:

1.Make sure that the tub has been cleaned, disinfected and rinsed BEFORE each use. Refer to Cleaning and Disinfecting on page 23.

2.Use the stretcher lift to transfer the resident to the tub.

The resident will remain on the stretcher lift during bathing.

3.Bathe the resident.

4.Drain the tub.

5.Transfer the resident from the tub.

6.Clean, disinfect and rinse the tub.

5.2Cleaning and Disinfecting

￿

￿

WARNINGS

Clean and disinfect the tub after EACH use to avoid resident infection and contamination of the tub. DO NOT clean or disinfect the tub when it is occupied. Otherwise injury may occur.

Read and understand ALL information on disinfecting BEFORE use.

ALWAYS wear rubber gloves, an apron and a face shield when using disinfectant. In case of eye or skin contact, follow procedures on the disinfectant container.

CAUTION

NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice stone. The use of these types of cleaners will make the gel-coat finish of your tub scratched and dull. NEVER use cleaners containing Iodine, Household Bleach, Bromine, Betadine® or Methylene Blue. The use of these types of cleaners will cause the tub to stain. Use of unapproved cleaners will dry out the rubber seals and gaskets and the tub will not function properly.

1.Use the drain plug to close the drain.

Part No. 1167456

23

Invacare® TheraPure™Supine Whirlpool Tubs

Page 23
Image 23
Invacare 6302G user manual 5.2Cleaning and Disinfecting, Warnings, Operation, Invacare TheraPureSupine Whirlpool Tubs