6 INSPECTION, MAINTENANCE AND TROUBLESHOOTING

6 Inspection, Maintenance and Troubleshooting

If any part of the system is not functioning properly, DO NOT use the system. For parts and technical service information please contact ICC Technical Service at (800) 668-2337, Extension 2.

6.1Safety Inspection Checklists

Every six months or as necessary have a qualified technician perform a thorough inspection and servicing. Routine maintenance will extend the life and efficiency of your tub.

Initially

Clean and Disinfect the tub. Refer to Cleaning and Disinfecting on page 23.

Visually inspect plumbing for leaks. Make repairs as necessary.

Ensure that the maximum water supply temperature does not exceed 110°F.

Ensure that the maximum water pressure does not exceed 55 PSI.

Verify the water flow temperature with a separate thermometer.

Visually inspect for cracks in the tub surface.

Check that all labels are present and legible. Replace if necessary.

Before Each Use

Clean and Disinfect the tub. Refer to Cleaning and Disinfecting on page 23.

Ensure that the maximum water fill temperature does not exceed 110°F (43° C).

Ensure that the maximum water pressure does not exceed 55 PSI.

Visually inspect for cracks in the tub surface.

Every Six Months

Verify the water flow temperature with a separate thermometer.

Visually inspect plumbing for leaks. Make repairs as necessary.

Visually inspect the power cord for damage or breakage. Replace if necessary.

Check that all labels are present and legible. Replace if necessary.

Clean the Water Temperature Control Valve to remove lime deposits. Refer to Removing Calcium Deposits, Scale and Lime Build-up on page 25.

6302GADH ONLY - Apply white lithium grease to Glide Pads. Refer to Refer to Maintenance for 6302GADH Adjustable Height Rub Panels and Glide Pads on page 30.

Part No. 1167456

29

Invacare® TheraPure™Supine Whirlpool Tubs

Page 29
Image 29