Replacing the Hydrophobic Filter
NOTE: For this procedure, refer to FIGURE 3.
NOTE: Replace the hydrophobic filter when either of the following conditions occur: (1) the filter is dirty and the aspirator has lost suction, or (2) there is liquid in the hydrophobic filter assembly due to extended use or overflow.
NOTE: The hydrophobic filter should be replaced every two months. However, if overflow occurs, change the filter immediately.
CAUTION
Use only Invacare recommended hydrophobic filters. DO NOT replace with other types of bacteria filters.
1.Remove the elbow connector end of the hydrophobic filter assembly from the top of the collection jar.
2.Separate the opposite end of the hydrophobic filter assembly from the vacuum gauge.
3.Disassemble the hydrophobic filter assembly by disconnecting the two connection tubes from the hydrophobic filter housing.
4.Discard the filter.
5.Re‐assemble the hydrophobic filter assembly with a new filter.
6.Re‐attach the hydrophobic filter assembly onto the collection bottle lid and vacuum gauge.
7.If necessary, twist the elbow connector to align with the collection jar.
| Hydrophobic |
Collection Bottle Lid | Filter |
Vacuum
Adjustment
Knob
Connection
ConnectionTubing
Tubing
FIGURE 3 Hydrophobic Filter Replacement
Part No 636959 | 7 | Aspirator Suction Unit |