7.8.1Error codes and diagnostic codes

The drive electronics are capable of rectifying some errors automatically. In this case the status display will cease to flash. Please switch the remote on and off several times. Wait approx. 5 seconds each time before switching the remote on again. If this does not rectify the error, locate the error using the flash codes shown below.

 

FLASH

 

 

FAULT

 

 

IMMEDIATE MEASURE

 

 

FURTHER HELP

 

 

 

 

 

 

 

 

 

 

CODE

 

 

 

 

 

 

 

 

 

 

 

1

 

 

Module defective.

 

-

 

 

Consult dealer.

 

 

2

 

 

Lifter raised or

 

If lifter is raised, lower in

 

-

 

 

 

 

 

 

lowered too far (seat

 

 

stages until the status

 

 

 

 

 

 

 

 

not at driving height)

 

 

display stops flashing. If

 

 

 

 

 

 

 

 

 

 

 

lowered too far, raise lifter

 

 

 

 

 

 

 

 

 

 

 

in stages until the status

 

 

 

 

 

 

 

 

 

 

 

display stops flashing. If at

 

 

 

 

 

 

 

 

 

 

 

all possible, only drive

 

 

 

 

 

 

 

 

 

 

 

when the seat is at driving

 

 

 

 

 

 

 

 

 

 

 

height.

 

 

 

 

 

 

 

 

Accessory error.

 

-

 

 

Consult dealer.

 

 

3

 

 

Error on right motor.

 

Check plug-in connectors.

 

Consult dealer.

 

 

 

 

 

Connection

 

 

 

 

 

 

 

 

 

 

 

loose/faulty.

 

 

 

 

 

 

 

 

4

 

 

Error on left motor.

 

Check plug-in connectors.

 

Consult dealer.

 

 

 

 

 

Connection

 

 

 

 

 

 

 

 

 

 

 

loose/faulty or motor

 

 

 

 

 

 

 

 

 

 

 

faulty.

 

 

 

 

 

 

 