SECTION 7—CYLINDER FILL TIMES

SECTION 7—CYLINDER FILL TIMES

NOTE: All filling times are approximate and may vary on environmental conditions.

*

Cylinder

Concentrator Flow Rate to Patient:

Filling

(Platinum 5 up to 3 L/min. and Platinum 10 up to 5

Times

L/min)

 

 

 

HF2PCM2KIT

 

 

NORMAL

25 min

 

 

MAXIMUM

30 min

 

 

Troubleshooting

If Conserving Cylinder fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provider.

PROBLEM

POSSIBLE CAUSE

 

SOLUTION

 

 

 

 

 

No flow.

1.

Regulator in “Off”

1.

Set to prescribed

 

position.

setting.

 

2.

Cylinder empty.

2.

Replace/fill cylinder.

 

3.

Dial not set at

3.

Set dial so indictor

 

increment.

points to a setting.

 

4.

Conserver not

4.

Check position of

 

sensing breath.

cannula in nose. Do not

 

 

 

breathe through mouth.

 

 

 

5.

Check to see if the

 

 

 

cannula is kinked.

 

 

 

 

 