Invacare MA80, MA85 manual Changing the Filter, Troubleshooting

Models: MA80 MA85

1 24
Download 24 pages 54.58 Kb
Page 23
Image 23

SECTION 3—MAINTENANCE AND TROUBLESHOOTING

Changing the Filter

NOTE: For this procedure, refer to FIGURE 3.1.

Clean the filter every 5 months or whenever it is dirty.

1.Remove the two screws securing the filter cover to the back of the control unit.

2.Remove the filter.

3.Wash the filter with water and a mild detergent.

4.Dry the filter completely.

5.Install the filter into the back of the control unit.

6.Secure the filter cover to the back of the control unit using the two screws.

Filter

Cover

Screws

FIGURE 3.1 Changing the Filter

Troubleshooting

PROBLEM

CAUSE

SOLUTION

 

 

 

Mattress not inflating

Mattress hose disconnected

Connect hose connectors and lock

Not alternating properly

 

them in place

 

 

 

 

Air hose kinked or split

Unkink hose or replace split hose

 

 

 

 

Major leak in the air cushions or

Replace leaking air cushions or

 

overlay pad

overlay pad

 

 

 

 

Kinked or split manifold

Unkink manifold or replace split

 

 

manifold

 

 

 

 

Has power and fuse is good, control

Send unit back for repair

 

unit does not come on

 

 

 

 

 

Not alternating, valve malfunction

Send control unit back to factory for

 

 

repair

 

 

 

 

No air, blower malfunction

Send unit for repair

 

 

 

No power

Control unit off

Check power source and turn on

 

 

unit

 

 

 

 

Power cord disconnected

Connect power cord to the power

 

 

source

 

 

 

 

No power in the power source

Check power source has power

 

 

power and turn it On

 

 

 

 

Power outage

Wait till the power source has

 

 

power

 

 

 

 

Blown fuse

Replace blown fuse with an

 

 

equivalent fuse

 

 

 

 

 

 

Part No 1148138

23

MA80 Series

Page 23
Image 23
Invacare MA80, MA85 manual Changing the Filter, Troubleshooting